Arthrex AR-8315 Manual Del Operador página 11

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 40
Explosion hazard-
Grounding-
Interfacing to equipment-
Cautions
Annual servicing-
Daily testing-
Performance-
Classification
Type of protection-
against electric shock
Degree of protection-
against electric shock
Mode of operation-
Do not use this equipment in the presence of flammable
anesthetics.
Do not defeat the three-wire grounding feature of the AC
adapter. A dangerous shock hazard may result.
Footswitches must be interfaced with other equipment. Be
certain to consult manufacturer's specifications to
maintain safe operation.
For continued safety and performance of the switch, it is
recommended that the functionality and electrical safety
of the switch be verified on an annual basis by an
authorized Arthrex Inc. representative.
It is essential that the footswitch be inspected everyday or
before use.
The infrared wireless footswitch system operates on a
wavelength of 870nm. It should be tested to assure
compatibility with any device it is connected to or
environment it is working in.
Report all problems experienced with the footswitch. If
the footswitch is not working properly, contact your
service representative for service. The footswitch should
not be used if it is not working properly.
Transmitter- Internally powered, no applied parts.
Receiver- Class 1 or 2 depending on power supply used.
Not applicable, no applied parts.
Continuous
P/N LM0659 Rev. 3
Page 11 of 281