Changing Vacuum Level; Placing Out Of Operation After Use - Medela Vario Instrucciones De Uso

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 290
Specifications:
Altitude
Vario 8
above
Max. Vacuum:
sea level:
low vacuum
+ 2000 m
– 7.0 kPa
+ 1000 m
– 7.9 kPa
+ 500 m
– 8.5 kPa
0 m
– 9.0 kPa
(Tolerance +/- 10 %)
7.3
Changing vacuum
level
7.4
Placing out of
operation after use
Vario 18 c/i
Vario 18
Max. Vacuum:
Max. Vacuum:
medium vacuum
high vacuum
– 43 kPa
– 59 kPa
– 48 kPa
– 66 kPa
– 51 kPa
– 70 kPa
– 55 kPa
– 75 kPa
1.
1.
or
2. If the pump is connected
to a fixed mains socket:
or
If the pump is connected
to a 12 V DC power source:
3. Clean and desinfect the Vario. See chapter 18.
1.1 Clamp patient tubing
1.2 Push and turn vacuum
regulator to select the cor-
rect vacuum according to
the particular application.
1.3 Check vacuum gauge for
setting.
1.1 Switch the on / off switch
of the Vario to position
«0» (c/i - versions: «Off»).
Disconnect the mains plug
from the fixed mains socket.
Disconnect the 12 V DC
plug from the 12 V DC
power source.
Operating instructions
| 17