Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 290

17 Filter test

1.
2.
3.
or
4.
1.1 remove tubing from filter.
2.1 Select maximum vacuum.
3.1 Switch on the Vario. The
pump will start running.
Caution: c / i versions
must be switched on in
«constant» mode.
4.1 read the vacuum.
4.2 replace the filter if
the vacuum exceeds
the values given in
the following table:
Vario 8
low vacuum
Vario 18
– 10 kPa
high vacuum
Vario 18 c / i
– 10 kPa
medium vacuum
Filter test
– 3 kPa
| 27

Publicidad

Tabla de contenido

Solución de problemas

loading

Tabla de contenido