Dentsply Sirona
Operating Instructions Suction devices for inLab MC X5 and
64 98 609 D3608
D3608.201.02.04.02 09.2017
2. Connect the cleaning hose to the upper opening of the suction
device.
3. Activate the manual operation for the suction device by setting the
switch at the top of the device from "Auto" to "On".
4. Suck up the dust from the chamber as required.
5. Sucking up as much dust as possible from the workpiece axis is
recommended. You must also always remove the dust from the
bottom of the impact pane of the block axis (see arrow).
6. Once the chamber has been cleaned using the suction device you
should complete the wet cleaning process lasting 30 seconds as
programmed in the software 2 to 3 times as required in order to
remove any accumulated material. This function is in the
configuration ..."
area of the CEREC SW / CEREC Premium SW
software if you click on the
cleaning process."
. Alternatively, you can also wet grind a normal
restoration with the same effect.
7. You can remove any zirconium oxide deposits in the lower part of the
workpiece axis with a nylon brush between the wet cleaning
processes.
6 Maintenance and repair
"Service"
button and then on
"Device
"Start a
21