6607290.book Seite 40 Donnerstag, 4. Januar 2018 11:55 11
6 Follow-up
40
Operating Instructions Midwest Stylus
6

Follow-up

6.1

After each treatment session

NOTICE! Condition immediately, or at the latest, one hour after
treatment.
NOTICE! Only use Midwest Plus Aerosol Spray.
✔ The bur must not be moving.
✔ Wear appropriate protective clothing.
1. Purge the water and air channels on the treatment center for 30
seconds.
2. Remove the bur with tweezers.
3. Predisinfect directly at the treatment center [ → 42].
4. Remove the high-speed handpiece.
5. Transport the high-speed handpiece to the hygiene room in a
suitable transport container.
®
®
Plus / Midwest Tradition
TC/Pro TCF
66 07 290 D3673.201.01.04.02