3. Ramp-Up Transition from
Stimulation to Expression
Phase.
When switching from Stimulation
to Expression Phase, the
Symphony Standard 2 .0 Card
increases the vacuum setting
by adding 3 bars to the last
Stimulation setting in a stair-stepped fashion . The
3 additional bars and increased vacuum setting
occur gradually over a 4-second period . The ramp-
up transition eliminates potential discomfort of an
immediate vacuum setting increase .
4. Use Let-Down Button to Switch
Back to Stimulation Phase.
The Symphony pump can switch
from Stimulation to Expression
Phase in one of two ways:
1) automatically after 2 minutes,
or 2) manually through pressing
the let-down button . With the
Symphony Standard 2 .0 Card, the let-down button
can also be used to switch back to Stimulation Phase .
For more information on this feature, please review the
Symphony 2-Phase Pumping Protocol information in
this booklet .
Note: Stimulation Phase is always limited to two minutes. If milk ejection
occurs before the 2 minute Stimulation Phase is over, press the let-down
button and continue pumping in Expression Phase
5. Program Remembers Last
Stimulation Vacuum Setting.
When switching back to
Stimulation Phase, the Symphony
Standard 2 .0 Card remembers
the last Stimulation Phase
vacuum setting prior to switching
to Expression Phase and returns to this setting in
Stimulation Phase .