· USE ONLY NSK genuine Tips when using NSK Varios Ultrasonic Scaler ( Varios 370 or Varios 370 Lux ) problems such as
damage, failure and accident of Handpieces resulting from use of Non-NSK Tips are not included in the warranty. The
following are the possible failure that could happen when using the Non-NSK Tips;
· Vibration failure caused by using non conforming screws.
· Patients accidental ingestion of broken Tips.
· Damage of thread ridge of handpiece.
· You must use the Tip within the power range described on the Tip-Power Guide. If you use it out of the power range, the
Tip might break or damage an operative site.
· When operating the product always consider the safety of the patient.
· Use by medical professional, such as doctor or dental hygienist, is intended.
· Check the vibration outside the patient's oral cavity before use. If any abnormalities are found, stop using immediately and
contact dealer.
· Do not drop, hit, or subject to excessive shock to the Control Unit/Handpiece.
· To prevent possible tooth plane damage and handpiece overheating, Always use with sufficient water.
· Do not sterilize by ultraviolet light. Handpiece could discolor.
· Sterilize the Tip, Handpiece,Tip Holder, Tip Cover S and Tip Wrench by autoclaving. Wipe the Control Unit,Tip Holder, Tip,
AC Adaptor, Foot Control, and Handpiece Cord including the cover.
· If chemical, solvent or antiseptic solution is deposited on this product, immediately wipe it away. Discoloration or
deformation may occur if left.
· Do not disassemble or alter the handpiece/Control Unit.
· Keep away from patients with cardiac pacemakers.
· Keep away from explosive substances and flammable materials. Do not use for patients anesthetized under laughter gas.
(Nitrous oxide)
· This product needs special precautions regarding EMC and needs to be installed and put into service according to the
EMC information.
· The use of ACCESSORIES, transducers and cables other than those specified, with the exception of transducers and
cables sold by the manufacturer of this product as replacement parts for internal components, may result in increased
EMISSIONS or decreased IMMUNITY of this product.
· This product should not be used adjacent to or stacked with other equipment and that if adjacent or stacked use is
necessary, this product should be observed to verify normal operation in the configuration in which it will be used.
· If any water drops remain on the handpiece or handpiece cord after autoclaving, wipe them off. Staining may result if left.
· There is the judgment that applies this product to a patient in the user side.
· During operation, high frequency oscillations in the handpiece and handpiece cord may affect computer and LAN Noise
may be heard during operation near a radio receiver.
· Be sure to turn off the Power/Volume Knob after use. Remove the AC Adaptor and water inside of the Control Unit before
storage.
· Users are responsible for operational control, maintenance and inspection.
· Clean/sterilize the product immediately after using it. Then store it. Leaving it non-sterile might lead to failure.
· When you have not used the product for long time and use it again, check the operation before use.
· Eye damage may result if the LED is stared directly into, Do not look into or turn it to the eyes of the patient.
· When abnormalities are found with a Control Unit and /or an AC adaptor, pull the AC adaptor from AC Outlet immediately.
· This product does not consider patient's age (except infants), gender, weight or nationality.
· No special training is required for this device.
· Applied parts for patient and/or operator are/is Tip and Handpiece.
* Operation Principle
A sinusoidal electrical signal, at ultrasonic frequency ( f > 20kHz ) , is delivered by the generator. This signal is applied
to the 'piezoelectric ceramic' located inside the transducer. Piezoelectric ceramic converts this signal into mechanical
vibrations. These vibrations are at the same ultrasonic frequency as the electrical signal. The mechanical vibrations
are propagated towards the distal end of the transducer. The "TIP" insert, which is attached at the distal end of the
transducer, vibrates at ultrasonic frequencies and makes it possible to achieve the aimed purpose.