  • ESPAÑOL, página 132
LED on
LED on bat­
power
tery charger
supply
Error after disconnecting the battery charger (error on self-test)
Beep
Error
signal
5 x long
Self-test error

14.2.3 Status signals

Battery charger is connected
LED on
LED on bat­
power
tery charger
supply
Battery charger disconnected
Beep
Event
signal
1 x short
Self-test completed successfully.
Battery charge level
The current battery charge level can only be displayed by connecting the battery charger and
power supply to the product.
Battery char­
ger
Battery charge level less than 50%
Battery charge level over 50%
Error
Battery is fully charged (or con­
nection with product is interrup­
ted).
Event
Power supply and battery charger operational
Resolution
is conducted which is confirmed
by beep signals.
When these beep signals are
heard, the battery is fully
charged.
If no beep signals are heard, the
connection to the product is
interrupted.
If the connection to the product is
interrupted, an authorised
Ottobock Service Centre must
inspect the product, battery char­
ger and power supply.
Resolution
Restart the product by connecting
and disconnecting the battery char­
ger (up to three times)
If the error persists, the product must
be
inspected
Ottobock Service Centre
by
an
authorised
65