Correct maintenance of the autoclave assures correct functioning of it and a secure saving
in terms of time and costs for assistance and maintenance.
11.1 Replacing the bacteriological filter
Substitute the bacteriological filter (PICT.2 pos. 11.) maximum every 200 cycles, or as
soon as the filter becomes a dark color.
Unscrew it in an anticlockwise direction (PICT.25). Screw on the new filter in a clockwise
direction (PICT.26) until it is seated.
Use only and exclusively original filters from the manufacturer of the autoclave.
11.2 Cleaning the chamber
Clean periodically the chamber, remove eventual deposits or debris, thus avoiding the
introduction in the discharging circuit of material which can cause obstructions.
For correct cleaning use only demineralized water and the abrasive sponge provided.
(PICT.13)
The filter in chamber is fixed about interference. When you estract the filter for the cleaning
please you don't damage it (PICT.27). Then,
control that protrude about 15mm from the chamber surface (PICT.29).
PICT.27
Never use solvents, detergents, chemical solutions, descaling agents or other similar
products.