24 Cleaning / Repair of the device
24.1 Cleaning
24.2 Repair service
24.3 Send in for repair
52
Danger of injury during maintenance, repair and transport
Danger from mains voltage.
Switch off the device and disconnect it from the power supply before
undertaking the cleaning work.
It is imperative that no moisture penetrate the inside of the
circulator.
Do not use alcohol- or solvent-based cleaning agents.
Have all service and repair work carried out by authorized electricians
only.
Regularly remove limescale deposits on the bath tank bottom with a
limescale remover. Clean the functional parts of the device with a soft
cloth and mild soapy water.
JULABO water baths are designed for continuous operation in normal
conditions. No regular maintenance work is required.
The bath tank should only be filled with suitable bath fluid. In the event
of contamination, the bath fluid must be replaced from time to time.
If Aqua Stabil is used, blue residues on the inside of the bath can be
removed with a cleaning agent based on anionic surfactants (e.g., dish
detergent). For this purpose, fill the bath with a diluted cleaning agent
solution and operate it for several hour at 60 °C.
Before calling out a service technician or sending back a JULABO device
for repair, we recommend that you contact our Technical Service
Department.
JULABO Technical Service
Phone:
+49 7823 5166
Fax:
+49 7823 5199
E-mail:
service.de@julabo.com
In the event that you send a device to JULABO:
Clean the device and, if necessary, detoxify it appropriately to avoid
the risk of harming the service personnel.