10.2 Limb Electrode Attachment

Electrodes should be placed on the soft skin of hands and feet. At first, clean the posi-
tions where limb electrodes are to be attached with alcohol, then applying conductive
cream to them.
Note: Screw tightly the knob of ECG cable's plug after it is inserted to the
ECG connector.

10.3 Check-List for Electrodes and ECG cables

Electrode Location
Right Arm
Left Arm
Left Leg
Right Leg
Chest 1
Chest 2
Chest 3
Chest 4
Chest 5
Chest 6
Cautions:
Do not turn on the device when installing lead cables.
If the ECG waveform for a long time does not appear, please check whether the elec-
trodes good contact with the skin.
Smear electrodes with conductive paste before installation.
Electrode Symbol
RA/R
LA/L
LL/F
RL/N
Vl/Cl
V2/C2
V3/C3
V4/C4
V5/C5
V6/C6
ENGLISH
Plug No.
9
10
11
14
12
1
2
3
4
5