Remove the caps from the rear hydraulic connections (water inlet, vent and drain water
tank
Intake (Pct. 3):
Connect the water intake hose (ref. CM61031, length 3 metres) to a ¾ G male valve.
Connect the other end of the hose to the ¾ G connection on the back of the cleaner.
Use the washers supplied (ref. CM50034) to make sure the connections are tight.
Discharge (Pct. 3):
Connect the drain hose (ref. CM61032, length 0.8 ‐ 2.7 metres).
FOR THE WATER TO DRAIN PROPERLY THE HOSE MUST DROP
BELOW THE CLEANER'S LEVEL.
Since the tank is emptied by gravity, the drain hose should not bend upwards to
prevent pressure drops that would considerably slow down draining and increase the
overall processing time.
Upstream from the outlet is a filter that preserves the operation of the drain solenoid
valve. Clean the filter regularly to keep the cleaner fully efficient at all times.
CLOSE THE VALVE UPSTREAM EVERY NIGHT OR WHENEVER
THE CLEANER REMAINS UNUSED FOR A NUMBER OF DAYS.
Hydraulic connections
A
B
C
vent/discharge outlet
D
E
Pct. 3
13
Filter
Discharge outlet (diameter 22 mm)
Detergent/disinfectant tank
Inlet connection (¾ G)
I/0 switch and power socket with fuses