Since there is no space between the electrode and the inner socket,
sweat cannot leak through
•
Easy to clean
•
Flexible electrode cushion compensates for changes in residual limb
volume
•
Proportional signal output – compatible with all Ottobock control units
•
Logarithmic sensitivity adjustment for exact patient fitting
•
Highly sensitive amplifier can use input signals as low as 5 µVeff for
digital controls and 15 µVeff for proportional controls
•
High common mode rejection in the low frequency range (>100 dB at 50
Hz) – providing rejection of electromagnetic interference, which may be
up to 100,000 times stronger than the input signal
•
High input resistance to compensate skin resistance differences
•
Electronic power-line frequency filter for additional interference protec
tion against signals with power-line frequency (40 dB suppression)
3 Application
3.1 Indication
The 13E202=* Suction Socket Electrode is to be used exclusively for the
control of
myoelectric prostheses of the upper limbs.
3.2 Application
The 13E202=* Suction Socket Electrode can be used for unilateral or
bilateral amputees.
3.3 Conditions of Use
The 13E202=* Suction Socket Electrode can be used for standard fittings
and in suction sockets.
3.4 Qualification
The fitting of patients with the 13E202=* Suction Socket Electrode
may only be carried out by a prosthetist.
4 Safety
4.1 Explanation of Warning Symbols
Warning regarding possible serious risks of accident or injury.
WARNING
Warning regarding possible risks of accident or injury.
CAUTION
16 | Ottobock