Optional Automatic Wash Cycle:
Any deviation from this guideline could result in damage to equipment as well as
improper cleaning results.
• Disassemble the device as indicated in section Cleaning of the MAYFIELD 2 Skull Clamp.
• Rinse components under warm tap water prior to placing in washer.
• Arrange device in a way to avoid contact between components.
• Select instrument cycle and ensure proper programming.
Phase
Pre-Wash 1
Enzyme Wash
Wash 1
Rinse 1
Thermal Rinse**
• Remove from washer and dry completely if needed.
• Inspect all components to ensure there is no visible organic debris or residue from cleaning
agent. Repeat process if any visible soil is detected.
(For additionally validated cleaning cycles, refer to Appendix A)
* No limit on tap water temperature
** Optional phase for disinfection of components - minimum water temperature as indicated or
per washer manufacturer specifications for the thermal rinse cycle
CAUTION!
Time in minutes
04:00
04:00
10:00
00:30
02:00
Table 3 Instrument Cycle