CAUTION
Incorrect assembly
Injury due to faulty operation or malfunction of the product.
►
Observe the following sections in these instructions for use.
CAUTION
Improper handling, assembly and disassembly
Injury due to faulty operation or malfunction of the product.
►
Instruct the patient in the proper disassembly, assembly and handling of the product.
INFORMATION
When the Protective Cover is installed, the battery charger cable has to point to the upper clos
ure. A correct knee joint charging process is only ensured with this alignment.
5 Scope of delivery
•
1 pc. 4X880=* Genium Protective Cover
•
1 pc. 4P880=U Upper closure
•
1 pc. 4P880=R Lengthwise closure
•
1 pc. 4P880=* Protector foot cuff
•
1 pc. 647G942 Instructions for use (qualified personnel)
•
1 pc. 646D298 Instructions for use (user)
6 Preparation for use
6.1 Required tools
•
718H5 Deburring knife
•
743B2 Tape measure
•
627B6=15 Crepe adhesive tape
•
636K13 adhesive (Loctite 241)
•
Vibrating/band saw
•
710D1 Torque wrench
6.2 Adapting the Protective Cover
6.2.1 Shortening the Protective Cover – Preliminary Shortening
1) Remove the prosthesis from the residual limb.
2) Remove the shoe from the prosthetic foot.
3) Loosen the foot connection cap of the prosthetic foot.
4) Check whether the closures on the Protective Cover are properly fastened and the Protective
Cover is not twisted.
5) Lay the Protective Cover next to the prosthesis.
INFORMATION: Make sure that the knee plateaus of the knee joint and Protective
Cover are at the same level (see fig. 1).
6) Mark the Protective Cover distally at the level of the foot contact surface (see fig. 2).
INFORMATION: For tall patients, shortening the Protective Cover may not be required.
7) Make another mark 1 cm above the existing mark.
8) Apply crepe adhesive tape proximally all around at this mark.
9) Use a vibrating or band saw to shorten the Protective Cover along the bottom edge of the
crepe adhesive tape.