Operating the Model 7500FO
Operating the Model 7500FO
NOTE:
Before using the Model 7500FO, please review all contraindications,
warnings and cautions.
Press the ON/STANDBY button. When the unit is first turned on, the
Model 7500FO performs a brief initialization sequence.
Verify that all LEDs illuminate and the unit beeps three times during the
first phase of the initialization sequence. If any LED is not lit (except the
AC Power Adapter LED), do not use the Model 7500FO. Contact NONIN
Customer Support for assistance.
To verify that the Model 7500FO is functioning properly, it is important to
monitor SpO
and pulse rate readings. Use the following procedure to
2
verify that the sensor is working properly.
1.
Ensure that the Model 7500FO is on, with the sensor connected.
2.
Apply the pulse oximeter sensor (see sensor instructions for use).
3.
Verify that a good SpO
appears, and that the pulse strength bargraph LED is active.
.
WARNING:
This device is intended only as an adjunct device in patient
assessment. It must be used in conjunction with other methods of
assessing clinical signs and symptoms.
.
WARNING:
As with all medical equipment, carefully route patient cables
and connections to reduce the possibility of entanglement or
strangulation.
reading is displayed, that a pulse rate value
2
11