Care and Maintenance

The advanced digital circuitry within the pulse oximeter of the
Model 7500FO requires no calibration or periodic maintenance other
than battery replacement by qualified technical professionals.
Field repair of the Model 7500FO circuitry is not possible. Do not attempt
to open the Model 7500FO case or repair the electronics. Opening the
case may damage the Model 7500FO and void the warranty. If the
Model 7500FO is not functioning properly, see "Troubleshooting."
CAUTION: Follow local, state and national governing ordinances and
recycling instructions regarding disposal or recycling of the device and
!
device components, including batteries. Use only NONIN-approved
battery packs. Batteries might leak or explode if used or disposed of
improperly.

Cleaning the Model 7500FO

Clean the Model 7500FO with a soft cloth dampened with isopropyl
alcohol, mild detergent, or a 10% bleach (5.25% sodium hypochlorite)
with water solution. Do not pour or spray any liquids onto the Model
7500FO, and do not allow any liquid to enter any openings in the device.
Allow the unit to dry thoroughly before reusing it.
.
WARNING:
Do not use this device in or around water or any other liquid,
with or without AC power.
CAUTION:
!
abrasive cleaning agents on the device. Do not gas sterilize or autoclave
this device. Do not place liquids on top of this device.
Clean the Model 7500FO separately from its associated sensors. For
instructions regarding cleaning pulse oximeter sensors, refer to the
appropriate pulse oximeter sensor package inserts.
20
Do not immerse this device in liquid, and do not use caustic or
Care and Maintenance