8-2 NBX MOTOR

The system incorporates an automatic diagnosis feature that can help to diagnose the cause of the problem
Details of each error code are listed below.
Error
Error Message
code
E–00
Over Motor current
Over PAM voltage
E–01
PAM diff. failure
E–02
Over EMF voltage
E–03
E–04
Torque diff. failure
Input voltage error
E–05
Over LED voltage
E–06
LED diff. failure
E–07
E–08
Over load
Motor start failure
E–09
E–11
Air FS failure
Comm. error
E–12
Over current(endo)
E–13
E–14
EEPROM error

8-3 SCALER

Error
Error Message
code
poorTool
E–09
E–14
Fatal
Cause
Excessive current is applied for a given
time at high speed mode.
Error in detecting the motor voltage
more than DC28V.
High speed Mode, Over difference
voltage of PAM control.
Low speed Mode, Over difference
voltage of EMF control.
Over difference voltage of Torque
control.
Input voltage has exceeded the upper
and lower limit values.
Detected over voltage of LED lamp.
Over difference voltage of LED lamp
control.
Motor is overused.
Handpiece/motor cord disconnection.
Air pressure is greater than 0.45[MPa]. Check user air pressure.
Disconnection or defective contact of
COM port.
Excessive current is applied for a given
time at low speed mode.
Memory read/write date error.
Cause
Handpiece cord disconnection.
Program error.
11
Solution
R e m o v e t h e l e a d f r o m m o t o r /
handpiece, and release foot control.
Release the foot control.
Release the foot control.
Release the foot control.
Contact your Authorized NSK Dealer.
Check input voltage.
Contact your Authorized NSK Dealer.
Contact your Authorized NSK Dealer.
Leave the motor until it cools down.
Check the handpiece cord is connected
with the handpiece properly.
Check the connection or wiring of COM
port.
R e m o v e t h e l e a d f r o m m o t o r /
handpiece, and release the foot control.
Contact your Authorized NSK Dealer.
Solution
Check the handpiece cord is connected
to the handpiece properly.
Contact your Authorized NSK Dealer.

Publicidad