ENGLISH
08
Take the cover away from the water tank. The cover is the punched plate with the knob.
Fill in the tank with distilled or demineralised water. Pour water until the lower edge of the
plate divider Photo B.
Attention: the water in the tank must not top the plate divider.
Put the cover on the water tank.
The use of feedwater with contaminants at levels exceeding those given
in the table above can greatly shorten the working life of the device, with
NOTE:
damage at the components of the autoclave and then invalidate the
manufacturer's warranty
8.1
Characterisstics of the water to be used
TABLE SHOWING THE QUALITY LEVELS LAID BY THE EN13060
Evaporation residue
Silicium oxide, SiO2
Iron
Cadmium
Lead
Rest of heavy metals except iron, cadmium, lead
Chloride (Cl')
Phosphate (P20s)
Conductivity (at 20°C)
PH value (degree of acidity)
Appearance
Hardness (E Ions of alkalin earth)
10
FILLING OF THE WATER TANK
Photo B
CEN STANDARD EN13060
10
1
0.2
0.005
0.05
0.1
2
0.5
15
5 to 7,5
Colourless clean without sediment
0.02
mg/l
mg/l
mg/l
mg/l
mg/l
mg/l
mg/l
mg/l
s/cm
mmol/l