Troubleshooting
Display Symbol
Condition/Cause
Measurement is below
normal temperature range.
Measurement is above
normal temperature range.
Operating temperature is
outside of the permissible
range 16.0°C – 35.0°C
(60.8°F – 95.0°F).
Blank display
Low battery
Low battery warning
Corrective Action
Retake temperature
ensuring proper
positioning.
Retake temperature
ensuring proper
positioning.
Take thermometer to a
room where the
temperature is within
the permissible range.
Wait 45 minutes to
stabilize thermometer.
Retake temperature
positioning the probe
snuggly into the ear canal.
Check battery polarity.
May need to replace
batteries.
Unit will work; replace
batteries soon.
Replace the batteries
immediately.