5.5
Connecting Insufflation Tube Set
Three different insufflation tube set types can be connected to the insufflation
tube connection at the front of the device (see Fig. 5-1 Device Front, page 16,
(3)).
Heated tube set with
Disposable (single use) insufflation tube set with:
Real-Time Pressure
• Filter
Sensing (RTP)
• Gas heating
• Measuring tube for direct pressure measurement
(RTP)
High Flow tube set
Disposable (single use) insufflation tube set with:
with Real-Time Pres-
• Filter
sure Sensing (RTP)
• Measuring tube for direct pressure measurement
(RTP)
High Flow II tube set
Disposable (single use) insufflation tube set with:
• Filter
Insert the plug of the insufflation tube set correctly into the insufflation tube set
connection at the front of the device until it snaps firmly into place.
• A short acoustic warning signal is emitted,
• the message Tube set connected is displayed,
• a check mark is displayed.
If the tube is accidentally unlatched by pressing twice or if the tube is removed
during insufflation:
• 2 long acoustic warning signal are emitted,
• the message Tube set not connected is displayed,
Operating the Device - General
Connecting the tube set
Set Pressure
Actual Pressure
mm Hg
mm Hg
ü
15
0.0
Liter
Tube set connected
Change
Actual
START
START
Mode
EN
Set Flow
l/min
3
0
l/min
19