9.3

Presetting Nominal Pressure in Vessel Harvest Operating Mode

The nominal pressure can be set during insufflation or while insufflation is
stopped. Values may range from 1 to max. 20 mm Hg or the value set in the Con-
figuration menu.
Increasing/decreasing nominal pressure:
Briefly press the º or » function field to increase or decrease the pressure. Keep-
ing the º or » function field depressed longer than 1.5 seconds activates scrolling
in increments of 1.
Safety limit:
When increasing the nominal pressure, the status line of the display depicts Safe-
ty limit starting at 12 mm Hg and the nominal value flashes.
The nominal pressure value 12 mm Hg is a limit value and should not be exceed-
ed when performing vessel harvesting surgery. Pressing the nominal pressure º
function field again does not increase the pressure any further.
CAUTION!
Exceeding this safety limit is to be decided by and the responsibility of the user/
operator.
Release the function field at that point. The display switches back to the nominal
value setting after 2 seconds. Now you can set a value up to 15 mm Hg.
The Safety limit message is re-displayed in the status line starting at 15 mm Hg
and the nominal value flashes.
The nominal pressure value 15 mm Hg is a limit value and should not be exceed-
ed when performing vessel harvesting surgery. Pressing the nominal pressure º
function field again does not increase the pressure any further.
CAUTION!
Exceeding this safety limit is to be decided by and the responsibility of the user/
operator.
Release the function field at that point. The display switches back to the nominal
value setting after 2 seconds. Now you can set a value up to 20 mm Hg.
9.4

Presetting Nominal Flow in Vessel Harvest Operating Mode

The nominal flow can be set during insufflation or while insufflation is stopped.
Increasing nominal gas flow:
• Press the function field º to increase gas flow. The nominal flow can be in-
creased from 1 to max. 10 l/min.
• Briefly press the corresponding function field to set values in increments of 0.5
for a range from 1 to 5 l/min.
• Set values in increments of 1 for a range from 5 to 10 l/min.
• Keeping the º or » function field pressed longer than 1.5 seconds activates
scrolling through the gas flow rates 1, 4, 10 l/min (in the Configuration menu).
Reducing nominal gas flow:
Using and Controlling the PNEUMO SURE XL High Flow Insufflator in Vessel Harvest Operating Mode
EN
53