Set up carrying case
The Invia Liberty is intended to be worn in a carrying case during operation.
1. Open the carrying case.
2. Slide the Invia Liberty into the designated pocket.
3. Close the velcro over the pump.
The shoulder strap can also be used as a belt strap.
To accomplish this, release the strap at the sides of the carrying case and pull it through the two loops
located on the back of the carrying case.
26
b) When the indicator is "half full" there is an air leak in the
system, but the pressure and therapy is maintained in
accordance with the set pressure.
➡
b)
c) When the indicator is "full" and flashes, there is a big air
leak in the system. When pump is first turned on and a big air
leak is present the air leakage alarm will go off within 2 minutes in
the Standard air leakage mode or within 5 minutes in the High air
leakage mode if the set pressure is not maintained. When therapy
is running and a big air leak is detected, alarm will go off within 2
➡
minutes. Follow the instructions shown on the display, or see the
c)
Alarms chapter.
Standby
Change the Invia Liberty pump into the Standby mode.
Press
"Standby"
running.
If the pump is in Standby mode for more than 5 minutes,
a warning will go off, follow instructions shown on the display
or see the Alarms chapter.
Turn OFF
Press [
> 3 seconds] and the pump will turn off.
[
> 3 seconds] and the pump motor will stop