INSTALLATION
!
The Turbo-V controller is equipped with a 3-wire power
cord and plug (internationally approved) for user's
safety. Use this power cord and plug in conjunction with
a properly grounded power socket to avoid electrical
shock.
High voltage developed in the controller can cause
severe injury or death. Before servicing the unit,
disconnect the input power cable.
The Turbo-V controller can be used as a bench unit or
a rack module, but it must be positioned so that free air
can flow through the holes. Do not install or use the
controller in an environment exposed to atmospheric
agents (rain, snow, ice), dust, aggressive gases, or in
explosive environments or those with a high fire risk.
During
operation,
the
conditions must be respected:
•
temperature: from 0 °C to +40 °C
•
relative humidity: 0 - 95%
(without condensation)
To connect the controller to the pump use the specific
cable supplied with the controller.
The controllers model 969-9405 and 969-9505 are
supplide with the pump cable fixed to the controller rear
panel.
See the appendix "Technical Information" for detailed
information about the above mentioned and the other
connections, and about the options installation.
USE
This paragraph describes the fundamental operating
procedures.
Detailed
procedures that involve optional connections or options
are supplied in the paragraph "USE" of the appendix
"Technical Information".
Make all vacuum manifold and electrical connections
and refer to Turbo-V pump instruction manual prior to
operating the Turbo-V controller.
WARNING!
NOTE
following
environmental
NOTE
information
and
operating
To avoid injury to personnel and damage to the
equipment, if the pump is laying on a table make sure it
is steady. Never operate the Turbo-V pump if the pump
inlet is not connected to the system or blanked off.
The input signal J1 connector should be left in position
including the shipping links if no external connections are
made. The forepump and the Turbo-V pump can be switched
on at the same time.
Controller controls, indicators and connectors
The following paragraph illustrates the Controller
control panel and interconnection panel. More details
are contained in the appendix "Technical Information".
1. Keyboard
push-button
selection. It is active only when the front panel
operation has been selected. Pressed once, the pump
runs at about 2/3 of the nominal speed. To unselect the
mode, press the push-button again.
2. Keyboard push-button for START, STOP, RESET
mode selection. It is active only when the front panel
operation has been selected. By pressing once the
starting phase begins; if pressed again it stops the
pump. If the pump has been stopped automatically by
a fault, this push-button must be pressed once to reset
the controller and a second time to restart the pump.
3. Keyboard push-button to recall on the display the cycle
number, cycle time and pump life.
4. Keyboard push-button to recall on the display the pump
current, pump temperature, pump power and rotational
speed. It is always active regardless of the operating
mode selected. Push-buttons 3 and 4, if pressed
together for at least 2 seconds, put the controller in a
routine where it is possible to program some operation
parameters.
5. LCD back-lighted alphanumeric display: dot matrix 2
lines x 16 characters.
Controller 969-9405,
969-9505 Front Panel
46
INSTRUCTIONS FOR USE
!
WARNING!
NOTE
for
LOW
SPEED
87-900-868-01 (C)
mode