Items to collect:
• PC with Hydrolab Operating Software
• Communications module
1. In the office or the field, connect the applicable communications module to a USB port on the PC.
2. Start the Hydrolab Operating Software. The connected communications module shows in the
Connect to Device field.
3. Select the communications module, then click Connect. The configuration windows for the
communications module shows.
4. Configure the communications module. Refer to Configuration in the online help.
5. When configuration is completed, disconnect the communications cable from the PC.
a. Select File>Disconnect from Device to disconnect the communications cable from the PC.
Note: As an alternative, click Devices, select the communications cable, then click Disconnect.
b. Disconnect the communications cable from the PC.
Connect the communications module
Items to collect:
• Data acquisition system or controller
• Communications module (configured)
• External power source, 6-24 VDC (12 VDC nominal)
1. In the field, connect the communications module to the data acquisition system or controller.
Refer to the documentation supplied with the communications module for wiring information.
Refer to
Data transmission and power wiring
transmission.
2. If the data acquisition system or controller cannot supply power to the communications module,
connect an external power source to the communications module. Refer to
power wiring
on page 17 for the wiring requirements for external power.
Data transmission and power wiring
To connect data transmission lines to the communications module, make sure that the transmission
cable is sufficient for the operating current and will transfer the data without distortion. For up to
305 m (1000 ft) of cable, three 26 AWG wires are sufficient.
To connect external power to the communications module, make sure to use two 18 AWG wires for
power. Smaller power wires can be used if the power supply is near the instrument.
Instrument deployment
Items to collect:
• Deployment cable
• Sensor guard
1. Connect the deployment cable to the communications module. Make sure to use the deployment
cable and not the calibration cable.
2. Connect the deployment cable to the instrument.
3. Remove the storage/calibration cup and install the sensor guard. Keep the storage/calibration cup
for later use. Refer to
16
Optional for instruments with an internal battery pack. Not used when the data acquisition
system or the controller can supply power to the communications module.
Figure 8
on page 14. Do not let the sensors become dry.
16
on page 17 for the wiring requirements for data
Data transmission and