3.2.2.63. Info.TitrResults.RS.1.Value
etc., up to .9
Info.TitrResults.EP.1.V
Info.TitrResults.EP.1.Meas
etc., up to .2
Info.TitrResults.Var.C40
etc., up to .C45
Info.TitrResults.Var.DTime
.RS: Values of the calculated results.
.EP: Endpoints:
Volume coordinate in mL, e.g. "1.2340"
Measured value coordinate in mV (with Ipol) "–241" or µA (with
Upol) "43.7".
.Var: Various variables. You may overwrite the variables C40...C45, see
3.2.2.62.
C40: Initial measured value in mV (with Ipol) "41" or µA (with
Upol).
C41: End volume in mL, "12.5360".
C42: Time from start of titration to end in s, "62".
C43: Volume drift on start of KFT titration from the conditioning in
µL/min, "3.5".
C44: Temperature in °C.
C45: Start volume in mL, "2.800".
DTime: Time for the drift correction in KFT with conditioning.
3.2.2.64. Info.StatisticsVal.ActN
Info.Statistics.1.Mean
Info.Statistics.1.Std
Info.Statistics.1.RelStd
etc. up to .9
The current values of the statistics calculation.
$Q sends, e.g.
ActN: Current value of the individual results
Data for MN1:
Mean: Mean value (decimal places as in result)
Std: Standard deviation (1 decimal place more than in result)
RelStd: Relative standard deviation (in %, 2 decimal places)
3.2.2.65. Info.SiloCalc.C24.Name
Info.SiloCalc.C24.Value
Info.SiloCalc.C24.Unit
for .C25 as for .C24
Info.SiloCalc.C26.ActN
Info.SiloCalc.C26.Mean
Info.SiloCalc.C26.Std
Info.SiloCalc.C26.RelStd
for .C27 as for .C26
The current values from the silo calculations. C26 is the mean value out of the
C24 variables; C27 comes from C25.
3.2 Remote control commands
read only/read+write
read only/read+write
read only
read only
read only
read only
read only
read only
read only
"3"
"3.421"
"0.0231"
"0.14"
read only
read only
read only
read only
read only
read only
read only