784 KFP Titrino
Selection of the values from the titrator which are sent in the set time interval
(see 3.2.2.88, formats see 3.2.2.70):
.CyclNo:
Cycle number. Together with the cycle time (see 3.2.2.75), a time
frame can be set up. The other data belong to the corresponding
cycle number. The cycle number is set to 0 at the start of a me-
thod and it is incremented until the end of the method.
.V:
Volume.
.dVdt:
associated volume drift.
.dMeasdt: associated measured value drift.
.dMeasdV: associated 1st derivative of the titration curve.
The unit "titrator" must be preset (see 3.2.2.90).
3.2.2.92. Setup.AutoInfo.Status
Setup.AutoInfo.P
Setup.AutoInfo.T.R
Setup.AutoInfo.T.G
Setup.AutoInfo.T.GC
Setup.AutoInfo.T.S
Setup.AutoInfo.T.B
Setup.AutoInfo.T.F
Setup.AutoInfo.T.E
Setup.AutoInfo.T.H
Setup.AutoInfo.T.C
Setup.AutoInfo.T.O
Setup.AutoInfo.T.N
Setup.AutoInfo.T.Re
Setup.AutoInfo.T.Si
Setup.AutoInfo.T.M
Setup.AutoInfo.T.EP
Setup.AutoInfo.T.RC
Setup.AutoInfo.C.B1
Setup.AutoInfo.C.R1
Setup.AutoInfo.C.B2
Setup.AutoInfo.C.R2
Setup.AutoInfo.I
Setup.AutoInfo.O
ON means that the Titrino reports automatically the moment the correspond-
ing change occurs.
.Status:
Global switch for all set AutoInfo.
.P
PowerOn: Simulation of power on (3.2.2.94). Not from mains.
Messages from node .T, Titrator:
.T.R
Ready: Status 'Ready' has been reached.
.T.G
Go: Instrument has been started.
.T.GC
GoCommand: Instrument has received a go command.
.T.S
Stop: Status 'Stop' has been reached.
.T.B
Begin of sequence.
.T.F
Final: End of determination, the final steps will be carried out.
.T.E
Error. Message together with error number, see page 49ff.
.T.H
Hold: Status 'Hold' has been reached.
.T.C
Continue: Continue after hold.
.T.O
Conditioning OK: EP reached (in KFT with conditioning).
3.2 Remote control commands
ON, OFF
ON, OFF
ON, OFF
ON, OFF
ON, OFF
ON, OFF
ON, OFF
ON, OFF
ON, OFF
ON, OFF
ON, OFF
ON, OFF
ON, OFF
ON, OFF
ON, OFF
ON, OFF
ON, OFF
ON, OFF
ON, OFF
ON, OFF
ON, OFF
ON, OFF
ON, OFF
ON, OFF