4
O
PERATION
4.1

Performing electrotherapy with the Vaco 200

1. Connect the Vaco 200 and the electrotherapy device to the mains
supply.
2. Connect the input connector A of the Vaco 200 to the output
connector A of the electrotherapy device with the connection cable.
If necessary, do the same for channel B.
3. Turn the vacuum intensity regulator to the
4. Use the switch on the rear panel to set the Vaco 200 and the
electrotherapy device to on. The LEDs from the traditional channels
of the Vaco 200 light up. The LEDs show which output channel is
connected with the electrotherapy device.
5. Select the desired electrotherapy program.
6. Connect, adjust and place the electrodes. See §4.2.1, §4.2.2, §4.2.3
and §4.2.4.
7. Turn the intensity knob A or B of the ET device to start the
electrotherapy and to set the desired intensity.
8. Check the patient's reaction. Repeat this check regularly during the
treatment.
9. The electrotherapy device stops the treatment and indicates that
the treatment is completed.
10. If vacuum is used, turn the vacuum intensity regulator to
safety, the vacuum stops automatically two minutes after the
treatment stops and the vacuum channel LEDs flash. Turn the
vacuum intensity regulator to
selection possible.
11. Remove the electrodes. If necessary put your finger under the rim of
the vacuum electrode for air inled.
Vaco 200
to make a new treatment
13
position.