Technical data
8
Performance data
Operating states,
features
Display
Probe connections
Accuracies
Measuring accuracies ∆pH ≤ 0,01 ±1 digit
Electrical Data
Measured variables
Battery
Current consumption
Simulation output
40
Technical data
Measurement pH with manual or automatic
temperature compensation (Pt 1000)
2-point calibration pH with buffer recognition and
possibility for adjustments
1-point calibration pH, zero point or slope
Measurement mV
Measurement °C
Simulation pH, mV
Manual temperature adjustment
8 mm LCD
Automatic polarity and simulation display, battery
control
Display of selected measuring mode, calibrating
functions and error messages
SN6 plugs
∆U ≤ 1 mV ±1 digit
∆T ≤ 0,5 °C ±1 digit
Measuring range
pH:
0,00 to14,00
mV:
-1300 to +1300
mV:
-199,9 to +199,9
°C:
-20,0 to +120,0
Temperature compensation of the pH value is only
made in the range of 0 °C to 100 °C
2 x 1.5 V alkaline batteries, size "AA", Mignon, IEC-LR6
max. 1.5 mA operating
≤ 50 µA standby
Output ranges according to the measuring ranges
pH and mV
Output resolution:
Absolute accuracy: better than 3 mV
Output impedance: 100 MΩ
Output setting time max. 60 seconds
Display resolution
0,01 pH
1 mV
0,1 mV
0,1 °C
1 mV