Cleaning; Contents; Ambient Operating Conditions - COOK Medical Rhapsody H-30 Manual Del Usuario

Ocultar thumbs Ver también para Rhapsody H-30:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Cleaning

Prior to cleaning the laser system, protect the fiber connection port with the fiber applicator
shutter closed. Unplug and power off the laser system prior to cleaning.
Do not use alcohol or an alcohol solution to clean the display.
Exterior enclosures may be wiped down with alcohol or another nonabrasive cleaner. Use a cloth
that has been sprayed with a cleaning or disinfecting agent. Avoid spraying the disinfectant
directly onto the laser system, particularly directly onto the touch screen. Do not let water
enter the laser system or leak into the fiber connection port. Solutions used for cleaning and
disinfecting should be allowed to evaporate before the laser system is used.

Contents

The Rhapsody H-30 Holmium Laser System is shipped with the following components and
accessories. Please take inventory of the items listed below and contact your designated Customer
Relations department if any items are missing or damaged.
Laser system
Foot switch
Bag that contains the following:
Service report form with final Quality Assurance checklist
Two keys for the key switch
Remote interlock connector
Two fill/drain tubes
Plastic water bottle for filling the cooling system
Vinyl cover over the laser system

Ambient Operating Conditions

The laser system should be used in a well-ventilated, air-conditioned operating room. The
procedure room temperature should be kept between 18 and 24°C (65-75°F) for optimum system
performance. In general, holmium lasers are very sensitive to temperature. When the laser cooling
system reaches above 38°C (100°F), the safety circuits will prevent the laser system from operating.
The information bar on the touch screen control panel will read "Over temperature. " The laser
system may overheat for the following reasons:
There is improper ventilation.
The ambient operating room temperature is above 25°C (77°F).
The laser system has been running at high power for an extended period of time (more than
30 minutes).
The input line voltage is lower than normal and has caused the cooling fans or water pump to
underperform.
Cook Medical © 2019
Return to Table of Contents
Return to Table of Contents
Return to Table of Contents
Chapter 4
Installation
1-23

Publicidad

Tabla de contenido

Solución de problemas

loading

Tabla de contenido