5.1 Install the Chiropro 3
FIG. 1
rd
A. Place the Chiropro 3
Gen on a flat surface capable of bearing
its weight.
CAUTION
It may be positioned on a table, on a trolley or any other surface
but in no circumstances on the floor.
FIG. 2
B. The fuse box may be opened with a screwdriver.
100 - 240 VAC = fuse T4.0AH 250 VAC REF 1307312-010.
To replace a fuse, see chapter"11.4 Replacement of fuses" on
page 23.
C. Connect the power cable (1) to the connector (2).
Note 1
CAUTION
The power plug is used for disconnection in case of problems, it
must be easily accessible at all times.
FIG. 3
D. Connect the pedal cable to the input provided on the rear panel,
guiding the connector and plug by means of the index pin on the
connector.
CAUTION
Do not lift the pedal holding the connection cable.
To disconnect the pedal cable pull the cable socket connector (1).
Do not pull the cable (2) without disconnecting the cable socket
before.
FIG. 4
rd
E. Connect the MX-i LED 3
output, guiding the connector and plug by means of the index pin on
the connector.
FIG. 5
F. Align and attach the bracket to the housing provided on the rear
of the console and suspend the flask or bottle.
FIG. 6
G. Check the packaging integrity, as well as the expiry date of the
irrigation line on the label (1).
CAUTION
The medical device must be used only with lines supplied by
Bien-Air Dental to ensure trouble-free operation. These lines are
sterile and for single use. Re-use may result in microbiological
contamination of the patient.
H. Remove the single-use sterile irrigation line (2) from its pouch.
rd
Gen system
Gen micromotor cable to the motor
FIG. 7
I. Connect the flexible hose of the irrigation line to the spray tube of
the handpiece or contra-angle.
FIG. 8
J. Install the peristaltic cassette (1) in the peristaltic pump (2).
Check that the cassette is clipped correctly.
FIG. 9
K. Close the pump lid (3). If there is resistance to closing, open the
lid again and check the correct positioning of the cassette. When the
lid is correctly closed, the user should hear a click sound.
CAUTION
Do not run the pump while the lid is open.
CAUTION
Do not run the pump without irrigation line.
CAUTION
Risk of pinching!
FIG. 10
L. Perforate the cap of the physiological liquid flask with the pointed
end of the irrigation line after removing the protective cap.
CAUTION
There is no detection of empty physiological liquid flask! Always
check the content of the flask before operating.
FIG. 11
M. Attach the irrigation line on the motor cable using the 3
attachment collars REF 1307727-010.
5.2 On/off procedure
The device can be switched on and off in complete safety using
the main switch on the Chiropro 3
CAUTION
Do not switch off the device while the motor is running.
NOTES
1
The equipment is powered by the mains power supply (100 -
240 VAC / 150W / 50-60Hz).
rd
Gen.
11