Bien-Air Dental Chiropro Plus 3rd Gen Instrucciones De Uso página 17

Ocultar thumbs Ver también para Chiropro Plus 3rd Gen:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 91
1
B. If necessary, release the footpedal to perform the following
actions:
Torquemeter (3) displays maximum reached value together
with the
symbol (4).
Torque bar (5) dots that were displayed in cyan turn black,
except for the maximum value dot which turns green.
• Turn the knob CW or CCW to respectively increase or
decrease the micromotor maximum reachable torque (quick
setting mode).
Note 12
The torquemeter (3) turns cyan and displays the set
micromotor maximum reachable torque.
Note 11
• Long press on the knob to change operation settings.
See chapter "9 Settings" on page 18 for details.
• Long press on the orange button to activate the 5 Ncm
torque boost.
Note 9 - 12
C. Short press on the footpedal's orange button or on the knob to go
to the next step.
The next step symbol turns green and the step's last used
settings are restored.
Note 7 - 10
FIG. 3
1
2
FIG. 4
NOTES
1
Real-time speed value is displayed in black when the
LED
micromotor is running. Maximum reachable speed value
stored is displayed in cyan when the
running, in steps P1 and P2.
2
Torquemeter is only displayed when micromotor speed is be-
low 100 RPM in steps P1 and P2.
3
The contra-angle ratio is cyan-colored for direct-drive and
green-colored for reduction gears.
4
Torque bar graph is only displayed when micromotor speed is
below 100 RPM, together with the maximum torque reached
during operation.
5
Each step settings are restored from the corresponding step
last used settings, excluding quick settings made directly in the
Operation mode.
6
In REVERSE mode, the rotation direction symbol
and there is a sound alert (alternate medium beeps). The torque
value is automatically increased in REVERSE mode when torque-
meter is displayed. The torque value can be increased from 0 to
10 Ncm, see chapter see " Reverse torque boost value" on page 20
to adjust it.
7
Actions on footpedal's buttons have no effect when the micro-
motor is running.
8
Changing the torque in steps P1 or P2 can only be performed
through the Settings mode.
9
The torque boost can only be activated when the torquemeter
is displayed in Operation mode, in low speed steps (<100 RPM).
10
For safety reasons, the speed setting icon turns red and
blinks together with the speedometer for 2 seconds when switch-
ing from low speed to high speed (≥100 RPM) step.
11
Changing the speed in steps P3, P4 and P5 can only be per-
formed through the settings mode.
12
Changes made in this mode (either by rotating the knob or by
changing parameters through the footpedal buttons) are consid-
ered as temporary settings and are never saved.
3
4
5
MX-i
MX-i LED
micromotor is not
blinks
15

Publicidad

Tabla de contenido
loading

Tabla de contenido