Weighing In Percent - Denver MAXX Serie Manual De Instrucciones

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Weighing in Percent

Purpose
This application program allows you to obtain weight readouts in percent
which are in proportion to a reference weight.
Example: Determine an unknown percentage; store the weight on the
balance/scale as the reference percentage (100%)
Step
1. Select application program
2. Select Weighing in Percent
3. Confirm setting
Symbol " " on the display:
application is active
4. Place empty container on the
balance/scale
5. Zero the balance/scale
6. Place the reference weight
for 100% on the balance/scale
7. If desired, change the number
of decimal places displayed: 100.0%,
100.00% or 100% (factory setting)
8. Confirm selected no. of decimal places
9. Place unknown weight on the
balance/scale
10. Toggle display between weight
and percentage
11. Unload the balance/scale
12. Weighing in Percent application:
clear the reference percentage
13. Reactivate Weighing in Percent
(if no other application program
has been selected)
14. Repeat procedure starting from Step 5
Key (or instruction)
(ZERO) ≥ 2 sec
(F) repeatedly
(ENTER)
(ZERO)
(F) repeatedly
(ENTER)
(F) repeatedly
(ENTER) ≥ 2 sec
(F)
.
Display