Press the PUMP CURRENT push-button and the
display shows (e.g.):
I
= 0
.
0 0 A
X
.
.
K R P M
where:
I = is the DC current drawn by the pump
range (0.00 to 5.00 Ampere)
P= is the DC power drawn by the pump
(range 0 to 999 Watt)
KRPM = is the theoretical rotational speed of
the pump as a function of the controller output
frequency (range 3 to 62 KRPM)
°C = is the temperature of the outer ring of the
upper bearing (range 00 to 99°C)
X = during operation a selected set point
condition (1 or 2 contrast inverted) appears
when the programmed threshold speed value
is not reached.
Press the CYCLE NUMBER push-button twice and the
display shows:
X X X X C Y C L E
P U M P
L I
P =
0 W
X X ° C
X X X X X m
F E
X X X X X h
where:
CYCLE = are the cycles performed (range 0 to
9999)
m = is the elapsed time related to the cycle
number displayed (range 0 to 99999 minutes)
PUMP LIFE = is the total operation time of the
pump (range 0 to 99999 hours).

Front / Remote / Serial Selection

Press
CYCLE
NUMBER
pushbuttons together for at least 2 seconds and the
processor enters in a routine where it is possible to
program the controller.
In this routine, the CYCLE push-button is used for
choosing/changing the value or condition;
the PUMP CURRENT push-button is used to enter and
confirm the value.
At any time it is possible to exit this routine by pressing the
CYCLE and PUMP CURRENT pushbuttons at the same
time for at least 2 seconds.
The display shows:
C O N F I
G U R A T I
S O F T W A R E
X X X
X X X X
and then:
S O F T
S T A R T
S E L E C T I
where: XXX = YES or NO.
If YES is selected, the Soft Start mode allows the pump to
rump-up the Normal speed within ten steps.
When NO is selected, the Soft Start mode is deselected
and the rump-up of the pump will be done within 150
seconds.
The controller is factory set to YES.
The Soft Start mode may be deselected/selected only
when the pump is stopped.
64
TECHNICAL INFORMATION
and
PUMP
CURRENT
O N
V E R S I
X X X X
M O D E
O N :
X X X
NOTE
87-900-893-01 (C)