Air Filter Maintenance; Troubleshooting - Apex XT Sense Manual De Instrucciones

Cpap system
Ocultar thumbs Ver también para XT Sense:
Tabla de contenido

Publicidad

FranÇais
4.4 Organigramme des paramètres de menu
Entrez le mode menu de l'utilisateur en appuyant sur la touche "MENU"..
Ramp
(Temps de la rampe)
Ramp P
(Pression initiale de la rampe)
P
(Pression de la thérapie)
CM
(Mesure de conformité)
Alarm (Alarme)
Clock Alarm (Réveil)
Clock Alarm <On>
Clock (Horloge)
Dans chaque réglage, lorsque chaque valeur de réglage a été modifiée, appuyez sur
"MENU"pour confirmer et appuyez sur "MENU" de nouveau pour la prochaine sélection de
réglage.
NOTE: Pour les médecins, veuillez vous référer à une page séparée intitulée
"Instructions supplémentaires à l'attention des médecins".
< 0, 5, 10, 15, ................, 45> min
< 3, 3.5, 4, 4.5, ................, 19> cmH2O
< 4, 4.5, 5, 5.5, ................, 20> cmH2O
< XXXX.X > hr
< on/off >
< hh: mm >
< hh: mm >
52
English

7.3 Air Filter Maintenance

For an optimum operation of the device, the air filter can be cleaned by the user at least once
every two weeks or more often if this device is operated in a dusty environment. It is
recommended that the filter be replaced with a new one after 6 months of use; however,
depending on the air quality, the replacement time may vary.
CAUTION:
Dirty air filter may cause high operating temperatures that affect device
performance. Ensure the air filter is cleaned and fitted at all times.
1. Remove the dirty filter from the enclosure on the rear of the
device.
2. Wash the filter in warm water with a mild detergent, and rinse
with water. Allow the filter to air dry completely before
reinstalling. Do not use a filter that is not completely dry. If the
filter is torn, replace it.
3. Reinstall the filter.
NOTE: Please follow national requirements to dispose the unit properly.

8. Troubleshooting

The table below lists troubleshooting solutions for the problems that may happen. If the problem
persists, contact your equipment provider service agent.
P r o b l e m
P o s s i b l e C a u s e s
No display
1. The power cord is not
connected to the power socket.
2. LCD failure or controlled PCB
failure.
Display code incorrect
LCD failure or controlled PCB
failure.
Illuminant under LCD is not
LED failure
on
Buttons disable
Button failure
Air delivered is slow
1.
During ramp time.
2.
Filter is too dirty.
3.
Flow generator failure.
Data can not be copied to
1. SD card is full.
the SD card
2. SD card is not inserted
correctly.
3. Data on the SD card is
corrupted.
S o l u t i o n s
1. Ensure the power cord is
connected.
2. Contact your equipment provider
for repair.
Contact your equipment provider for
repair.
Contact your equipment provider for
repair.
Contact your equipment provider for
repair.
1. Check the ramp time setting
2. Change or clean the filter
regularly.
3. Contact your equipment provider
for repair.
1. Ensure the SD card has enough
capacity.
2. Ensure the SD card is inserted into
the slot.
3. Format the card.
13