5. Press the power switch. The compressor
turns off and nebulization stops.
6. Disconnect the air tube from the nebulizer
kit. Hold the air tube plug and gently pull down to disconnect the air tube
plug from the air tube connector on the bottom of the nebulizer kit.
7. Check the air tube. No condensation or moisture should remain in
the air tube.
WARNING
Do not store the air tube with moisture or medication remaining in
the air tube.This could result in infection as a result of bacteria.
If any condensation or moisture remains in the air tube, remove the moisture
from the air tube. Follow the directions below:
1. Make sure the air tube is still connected to the air tube connector on
the front side of the compressor.
2. Press the power switch on. The compressor will start and pump
air through the air tube to expel the moisture.
3. Press the power switch again to turn the compressor off.
8. Disconnect the air tube from the compressor. Hold the air tube plug and
gently pull the air tube plug off the air tube connector on the front side
of the compressor.
9. Unplug the device from the electrical outlet.
CAUTION
Unplug the power cord from the
electrical outlet after using the device.
CAUTION
Remove the rechargeable battery and the
AC Adaptor from the device after use.
USING THE DEVICE
16
Power Indicator