The screen will show:
If you select "Yes", then the heated lid will work (holding
at 105 ºC) when running. You don't need to add oil in the
tube.
IMPORTANT: Attention please, BE SURE NOT TO ADD
OIL IN THE WELL OF BLOCK AT ANY TIME.
If you select "No", then you must add oil in the tube,
otherwise the sample in the tube will be condensed at the
top of the tube.
If you select Sim-Tube control mode in your protocol, the
screen will show at this time:
After you select the tube or plate you use, the screen will
show:
Input the volume of your sample, then the instrument will initiate running program. If
you turn-on the heated-lid, the lid will firstly enter pre-heating stage. When the heated lid
reaches 100 ºC, the instrument begins running program. The heated lid continues heating
until it reaches 105 ºC.
Whenever the program finishes normally or abnormally,
the heated lid will shut off automatically. In addition, you
could select LID in main menu to set the temperature that
the heated lid turn-off automatically.
If you select turn-off the heated lid, we have to remind
you once again that you must add oil in the tube, at this time the protocol will be laun-
ched without lid Heating display.
When running, you could see the normal running display from the LCD window:
RUN:
a1
Hold 40C Blk=39.9C
0m30s
¡Step 2
If the step is a gradient step (only 4196 model),
you could also see an additional gradient informa-
tion on the normal running display:
Page 28
Instruction manual 50604xxx
ENGLISH
RUN:
A1
a2
a3
a4
a5
a6
RUN:
a1
Use heated lid?
YES
no
RUN:
a1
Vessel:
0.2 TUBE
0.5 tube
plate
RUN:
a1
Volume(ul):
0
RUN:
a1
Lid temperature : 20 ºC
RUN:
a1
Grad
30C
50C
Hold
40C
Blk=39.9C
0m30s
Step 2
Version 1 April-08
Pulse [ENTER], visualizará
Step # 1
Option? NO yes
Funciones opcionales:
1. Incremento / disminución temperatura
2. Incremento / disminución tiempo
3. Ajuste de la tasa de incremento
4. Establecimiento de la Auto - Pausa
5. Función de gradiente (Modelo 4196)
Si no precisa de ninguna de las funciones opcionales de Option, mueva el cursor hasta
No, pulse [ENTER] para pasar al siguiente paso..
Si precisa de alguna, mueva el cursor a Yes, presione [ENTER] para mostrar una pági-
na de opciones.
Step # 1
INC ext ramp pause
Enter:
a1
Step # 1
Si quiere seleccionar Incremento / Disminución de la Temperatura,
mueva el cursor a Inc, y luego presione [ENTER].
Delta temp. = 0.0C
Nota: La Opción de Incremento / Disminución de la temperatura puede ser usado en un
experimento de PCR tipo Touchdown. El rango de temperatura es de -10 ºC a 10 ºC.
Si precisa de un incremento de temperatura de 0.5
de una disminución de la temperatura de 1 ºC, presione [-] [1] [0]. Pulse
[ENTER].
Si necesita Incrementar/ Disminuir el tiempo, mueva el cursor a Ext y presione
[ENTER].
Enter:
a1
Step # 1
Delta time = 0s
Nota: El Incremento / Disminución de Tiempo puede ser utilizado en un experimento de
tipo Long – PCR.
Si quiere seleccionar un incremento de Tiempo de 1 segundo, presione [1] y [ENTER].
Si necesita de una disminución del tiempo en 1 segundo, presione [-][1] y [ENTER]. Si el
display es correcto, pulse [ENTER] para volver a la página de opciones. Pulse [▼] [▲] o
[STOP], en la pantalla verá:
Step # 1
Option? no YES
Revisión 1, Abril-08
Manual de instrucciones 50604xxx
CASTELLANO
º
C, presione [5]. Si requiere