3. S
TERILISATION AND
3.5.2 Notes
a) [Note 1]
The water control knob can adjust the water volume according to the symbol.
b) [Note 2]
Clean the water pipe with the multi-function syringe of the dental unit (as showed).
Multi-function syringe
Cut the water pipe at a distance of 10cm~20cm from the water entrance.
Turn on the power switch.
Connect the multi-function syringe of dental unit to the water pipe.
Disassemble the tip or handpiece.
Step on the foot switch.
Turn on the switch of the multi-function syringe,press the water into the machine and the
impurity blocked in the water pipe can be eliminated.
c) [Note 3]
If the scaling tip has been screwed on tightly and there is fine spray too, the following phe-
nomena show that the scaling tip is damaged:
• The vibrating intens ity and the w ater atomization degree become weak obviously.
• During treatment, it produces the sound like buzzfrom the scaling tip.
M
V
P
S
PS1
AINTENANCE
ET
IEZO
CALER
12
4.P
RECAUTION

4.1 Notice when using equipment

4.1.1 Keep the scaler clean before and after operation.
4.1.2 The handpiece, scaling tip, torque wrench, endo wrench and endochuck
must be sterilized before each treatment.
4.1.3 Don't screw the scaling tip and endochuck when stepping on the foot
switch.
4.1.4 The scaling tip must be fastened and there must be fine spray or drip
coming out form the tip when operating.
4.1.5 Change by a new one when the tip and ultrasonic file are damaged or
worn excessively.
4.1.6 Don't twist or rub the tip and endochuck.
4.1.7 Don't use impure water source and be sure not use normal brine
instead of pure water source.
4.1.8 If use the water source without hydraulic pressure, the water surface
should be one meter higher than the head of the patient.
4.1.9 Keep the connector of handpiece and the socket of the cable dry before
installing the handpiece.
4.1.10 Don't pull the cable forcibly in case of the handpiece falling off from
the cable.
4.1.11 Don't knock or rub the handpiece.
4.1.12 After operation, turn off the power, then pull out the plug.
4.1.13 We are only responsible for the safety on the following conditions:
The maintenance, repair and modification are made by the manufacturer or
the authorized dealer.The exchanged components are original of HENRY
SCHEIN INC. and operated according to instruction manual.
4.1.14 The internal screw thread of the scaling tips produced by some man-
ufacturers maybe coarse, rusty and collapsed.
This will damage the external screw thread of the handpiece irretrievably.
Please use HENRY SCHEIN brand scaling tips.
V
P
S
PS1
ET
IEZO
CALER
13