2.7.3.
POWERCELL MTX
2.7.3.1.
Jumper Settings
Two jumpers on the POWERCELL MTX board are designed W1 and W2. If the application contains
more than 14 POWERCELLs, an external Power Supply (P/N 0917-0240) must be connected on
pins 9 and 10 of the board connector, and jumper W2 should be in the 24V position. Figure 2-45
shows the jumper's locations and settings. Table 2-9 describes the purpose of each setting.
2-28
METTLER TOLEDO IND780 Installation Guide
W2 (Pin 1)
W1
Figure 2-45: POWERCELL MTX Card Jumper Locations and Settings
Table 2-9: POWERCELL MTX Jumper Description
Jumper
On
Off
This jumper should be closed when a network terminating resistor is required
X
to be installed in the terminal. This resistor is almost always required.
W1
This jumper should be open when a network terminating resistor is not to be
X
installed in the terminal.
Pins
When 14 or fewer load cells are connected, internal power (12V) can be
1, 2
used, and the jumper should close pins 1 and 2.
W2
When more than 14 load cells are used, external power is necessary and the
2, 3
jumper should close pins 2 and 3.
Figure 2-44: POWERCELL MTX Option Board
Pin 1
Voltage and Terminating
Resistor Setting Jumpers
Description
64057253
|
14
|
08/2014