USB CONTROL
USB Interface
Remote Functions:
Setting Accuracy:
Setting Resolution:
Readback Accuracy:
Readback Resolution:
Remote command
processing time:
GENERAL
AC Input:
Power Consumption:
Operating Range:
Storage Range:
Environmental:
Safety & EMC:
Size:
Weight:
Standard, ground-connected, USB 2.0 hardware connection (opto-
isolated from power supply output). Operates as a virtual-COM port.
Set voltage, Set Current, Set Output On/Off, Read Voltage, Read
Current, Read On/Off, Read Mode (CV or CC).
Voltage 0.3% ± 20mV (for V >0.1V).
Current 0.6% ± 20mA (for I >0.1A).
Voltage 10mV. Current 10mA.
Voltage 0.3% ± 100mV. Current 0.6% ± 20mA.
Voltage 100mV. Current 10mA.
Typically <10ms between receiving the command terminator for a step
voltage change at the instrument and the output voltage beginning to
change.
220V-240V AC ± 10%, 50/60Hz.
110V-240V AC ± 10% by factory option.
Installation Category II.
400VA max.
+5ºC to +40ºC, 20% to 80% RH.
−40ºC to + 70ºC.
Indoor use at altitudes up to 2000m, Pollution Degree 2.
Complies with EN61010-1 & EN61326-1.
For details, request the EU Declaration of Conformity for this
instrument via http://www.aimtti.com/support (serial no. needed).
140 x 160 x 295mm (WxHxD).
2.8 kg
3