InSUS Head for Optical Single-Use O
4
Operation
4.1 Calibration
Calibration must be performed every time a new single-use
sensor is connected to the InSUS Head via the InSUS optical
fiber cable. For most applications, a single point calibration
should be sufficient. The calibration medium can be either
air or a calibration gas with known O
medium with a known O
calibration in gas, the correct pressure and the correct hu-
midity have to be set in the transmitter. For detailed informa-
tion please refer to the transmitter or biocontroller manual.
5
Maintenance
5.1 Visual inspection
To check your sensor head, we recommend the following
procedure:
– The contacts of the connector must be dry: moisture, cor-
rosion and dirt in the connector can lead to false readings.
– Check the cable for buckling, brittle areas or ruptures.
– Check the contacts of the InSUS fiber optical cable.
6
Environmental protection
Waste electrical products should not be dis-
posed of with household waste. Please recycle
where facilities exist. Check with your Local
Authority or retailer for recycling advice.
7
Warranty
On manufacturing defects, 12 months after delivery.
8
Storage conditions
Please inspect the instrument immediately after receiving for
any transport damage. The sensor Head should be stored in
a dry, clean and protected area until the time of installation.
If the sensor Head has been removed from the process, it
should be thoroughly cleaned and dried. It should be stored
in a dry, clean and protected area until time of installation.
9
EC Declaration of Conformity
The complete declaration is available in the declaration of
conformity certificate.
InTrac is a registered trademarks of the METTLER TOLEDO Group
in Switzer l and, the USA, the European Union and a further five
countries.
InSUS is a trademark of the METTLER TOLEDO Group.
© 12 / 2020 METTLER TOLEDO
Printed in Switzerland
Sensors
2
concentration or liquid
2
concentration. Before starting the
2
en
13
InSUS Head
30 313 265 B