facturer of the appliance may cause an increase in the electromagnetic emissions and/or a
reduction in the electromagnetic immunity of the device, thus causing a malfunction.
Portable and mobile radio communication devices (mobile phones, transceivers, in-
cluding peripheral devices like antennas cables and external antennas, etc.) may affect the
medical device and should not be used close to (at more than 30cm from any part of the
device, including cables), adjacent to or overlapped on the medical device. If such use is
necessary and inevitable, special precautions shall be adopted so that the electromedical
device works properly in its standard configuration (for instance, by steadily and visually
checking the absence of anomalies or malfunctions).
The tables below provide information on the EMC characteristics (Electromagnetic Compat-
ibility) of this electromedical appliance.
Guide and declaration of the manufacturer – Electromagnetic Emissions
The Mistral aerosol can be used in the following electromagnetic environment. The Cus-
tomer and/or the user of the Mistral aerosol shall make sure that the appliance is used in
such environment.
Emission test
Radiated emissions /
Conductions CISPR11
Radiated emissions /
Conductions CISPR11
Harmonic currents EN
61000-3-2
Voltage fluctuations /
flicker EN 61000-3-3
Guide and declaration of the manufacturer – Electromagnetic Immunity
The Mistral aerosol can be used in the following electromagnetic environment. The Cus-
tomer and/or the user of the Mistral aerosol shall make sure that the appliance is used in
such environment.
Immunity test
Electrostatic
discharge (ESD)
EN 61000-4-2
Fast transient / burst
EN 61000-4-4
Conformity
Group 1
Class [B]
Class [A]
Compliant
Standard of proof
± 8kV contact
± 15kV air
± 2kV power supply
± 1kV signal cables
ENGLISH
Guide to the electromagnetic
environment
The Mistral aerosol uses RF energy only for
Internal function. Therefore, its RF emissions
are very low and do not cause any interfer-
ence with other nearby electronic applianc-
es.
The Mistral aerosol is designed for use in
any environment, including households and
those directly connected to the public power
distribution grid which supplies power to en-
vironments intended for domestic use.
Level of
Guide to the electromag-
conformity
netic environment
The appliance
Floors should be in wood,
does not alter
cement or ceramics. If floors
its status
are covered by synthetic
material, the relative humid-
ity should be at least 30%.
The appliance
The power supply should be
does not alter
that of a typical commercial
its status
premise or hospital.