NSK VIVA MATE G5 Manual De Instrucciones página 23

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 69
Guidance and manufacturer's declaration - Electromagnetic Immunity
VIVA MATE G5 is intended for use in the electromagnetic environment specified below. The customer or the user of VIVA MATE G5 should assure that it is used in
such an environment.
Immunity test
IEC60601 test level
Conducted RF
3Vrms
EN/IEC61000-4-6
150kHz to 80MHz
Radiated RF
3V/m
EN/IEC61000-4-3
80MHz to 2.5GHz
NOTE1: At 80MHz and 800MHz, the higher frequency range applies.
NOTE2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
a: Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio
broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an
electromagnetic site survey should be considered. If the measured field strength in the location in which the product is used exceeds the applicable RF compliance
level stated above, the product should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such
as reorienting or relocating the product.
b: Over the 150kHz to 80MHz frequency range, the field strength should be less than 3V/m.
c: Externally displayed symbol of a device or component of device comprising an RF transmitter or emitting RF magnetic field energy for diagnosis or
treatment.
Cable and Accessories
Motor Cord
Foot Control Cord
AC Adaptor
Recommended separation distances between portable and mobile RF communications equipment and VIVA MATE G5
VIVA MATE G5 is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of VIVA MATE G5
can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and
VIVA MATE G5 as recommended below, according to the maximum output power of the communications equipment.
Rated maximum output power of
transmitter
W
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation distance 'd ' in meters (m) can be estimated using the equation
applicable to the frequency of the transmitter, where 'P' is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE1: At 80MHz and 800MHz, the higher frequency range applies.
NOTE2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
Compliance level
3 Vrms
3 V/m
Max length
Complies with
1.5m
RF emission CISPR11
2.0m
1.8m
Electrostatics discharge(ESD)
Surge immunity
Power frequency magnetic field
Conducted RF
Radiated RF
150kHz - 80MHz
d =1.2 P
0.12
0.38
1.2
3.8
12
Electromagnetic environment - guidance
Portable and mobile RF communications equipment should be used
no closer to any part of the product, including cables, than the
recommended separation distance calculated from the equation
applicable to the frequency of the transmitter.
Recommended separation distance
d = 1.11 P
d = 1.00 P 80MHz to 800MHz
d = 2.00 P 800MHz to 2.5GHz
Where P is the maximum output power rating of the transmitter in
watts (W) according to the transmitter manufacturer, and (d) is the
recommended separation distance in meters (m).
Field strengths from fixed RF transmitters as determined by an
electromagnetic site survey
level in each frequency range
Interference may occur in the vicinity of equipment
marked with the following symbol:
nonionizing radiation (IEC60417-1:2002 5140)
Separation distance according to frequency of transmitter
m
80MHz - 800MHz
d =1.2 P
0.12
0.38
1.2
3.8
12
(a)
should be less than the compliance
(b)
.
Class B/ Group 1
EN/IEC61000-4-2
EN/IEC61000-4-5
EN/IEC61000-4-8
EN/IEC61000-4-6
EN/IEC61000-4-3
800MHz - 2.5GHz
d =2.3 P
0.23
0.73
2.3
7.3
23

Publicidad

Tabla de contenido
loading

Tabla de contenido