Incident Description
A critical measurement error is
detected on internal voltage reference
A critical measurement error is
detected on internal voltage reference
A critical measurement error is
detected on Input 1
A critical measurement error is
detected on Input 2
A critical measurement error is
detected on the ground measurement
The device is in simulation mode and
one or more of its Device Variables are
not representative of the process
Description
LED reaction
Drift error, reference
Flashing Red
voltage FVR
Drift error, reference
Flashing Red
voltage VREF
Drift error, primary
Flashing Red
Input 1
Drift error, primary
Flashing Red
Input 2
Drift error, ground
Flashing Red
voltage offset to
terminal 3
Device Variable
No Impact
Simulation Active
5437V101-ES
Analog Output
NE-107 Class
Reaction
Safe State
Failure
Safe State
Failure
Safe State
Failure
Safe State
Failure
Safe State
Failure
No impact
N.A.
User action
Reconnect or repair sensor.
If the error is persistant send in the
device for repair
Reconnect or repair sensor.
If the error is persistant send in the
device for repair
Reconnect or repair sensor.
If the error is persistant send in the
device for repair
Reconnect or repair sensor.
If the error is persistant send in the
device for repair
Reconnect or repair sensor.
If the error is persistant send in the
device for repair
N.A.
Error
#
41
42
43
44
45
46
53