We recommend using the disinfecting wipes/towelettes below for
meter cleaning and disinfection because the active ingredients have
been tested to be effective against Hepatitis B Virus (HBV) for FORA
blood glucose meter.
Micro-Kill+™ (Micro-Kill Plus™) by Medline (EPA Reg. No.
59894-10-37549)
To obtain disinfecting wipes and other information, please contact
Medline at 1-800-MEDLINE (1-800-633-5463) or visit www.medline.
com. You can also purchase at www.amazon.com.
Disinfecting Procedures
1. Take out one disinfecting wipe from the package and squeeze
out any excess liquid in order to prevent damage to the meter.
2. Wipe all meter's exterior surface display and buttons. Hold the
meter with the test strip slot pointing down and wipe the area
around the test slot but be careful not to allow excess liquid to
get inside. Keep meter wet with disinfection solution contained
in the wipe for a minimum of 2 minutes for Micro-Kill+™ wipes.
3. Remove the wipe. Allow the meter surface to dry completely.
4. Discard the used wipes and never reuse them. Wash your hands
thoroughly with soap and water after handling the meter, lancing
device and test strips to avoid contamination.
This device has been validated to withstand up to 5,000 cleaning
and disinfection cycles using the recommended disinfecting wipe/
towelette. The tested number of cycles is estimated by 5 cleaning