Various Functions - NSK VarioSurg 3 Manual De Operación

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 73
Use the Tip ONLY in the Mode, and within the Power Range, as
displayed on the Tip case label. If used outside the prescribed
CAUTION
power range the Tip may break, and the tooth surface or soft
tissue may be damaged. (Fig. 21)
ONLY tips specified for maintaining prostheses can be applied
to prostheses. Contact of other types of tips with prostheses
may cause Tip breakage.
7-3 Activation
The Tip will commence vibrating when the ON-OFF Foot Control button is depressed.
Start with a low power setting within the power range, observing the patient's reaction and according to
the treatment plan.
CAUTION
Always use the irrigation. If the irrigation is not sufficient, the Handpiece may excessively heat, or oral
injury may occur.
Do not apply excessive pressure to the Tip.
Do not place the active Tip anywhere other than the specific treatment area, as the ultrasonic vibration
may cause oral injury.
Even if the power level is correctly set for the Tip in use application of excessive pressure to the Tip may
cause Tip damage. Should Tip breakage occur, immediately evacuate the oral cavity of any Tip pieces
using strong suction.
If Tip vibration reduces during operation, release the Foot Control activation button, and then again
depress the Foot Control activation button. If the vibration level still does not return to normal check that
the Tip is correctly and firmly positioned. Incorrectly inserted Tips may loosen during operation and power
loss may be experienced.
Do not connect or disconnect the AC power or the Handpiece Cord while Foot Control is depressed.
8

Various functions

8-1 Memory function
The set values displayed in the panel (Eg. Mode, Program No., Power Level, Irrigation Flow Level, illumination intensity)
can be memorized in each program. Once memorized, the programs can be called up whenever required.
1) Press the PROGRAM key and select the Program No. for memory.
2) Check the values of the parameters to be stored.
3) Hold down the MEMORY key for approximately one second. When the notification tone beeps, memorizing the
setting is complete.
8-2 Volume selection function for the notification tone
Sound volume of the notification tone can be selected between high & low.
1) Make sure the power is off. Then turn ON the power while holding down the BURST key.
2) The notification tone in the selected volume sounds.
3) To change the sound volume, repeat 1) and 2) above.
8-3 Link function with Surgic Pro
This product has a link function whereby two systems (NSK VarioSurg 3 and NSK Surgic Pro) can be operated via one
Foot Control.
* To use this function, Surgic Pro (sold separately) and the SG Link Set (sold separately) are required. For the Link
function, refer to the Operation Manual included with the SG Link Set.
Mode
14
SG5
Z305106
POWER RANGE
S
Power range
Fig. 21

Hide quick links:

Publicidad

Tabla de contenido

Solución de problemas

loading

Tabla de contenido