NSK VarioSurg 3 Manual De Operación página 17

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 73
9-3 Cleaning the Glass Rod
Wipe clean the Glass Rod tip with an alcohol immersed cotton swab.
Remove all debris. (Fig. 23)
Do not use needles or blades when cleaning the
Glass Rod as the glass surface may be scratched and
CAUTION
illumination intensity will be reduced.
9-4 Cleaning the Control Unit and the Foot Control
1) Turn OFF the power and disconnect the AC power cord from the wall power socket.
2) Wipe the surface, first with a moist cloth, then with an alcohol soaked cloth.
9-5 Cleaning other components
Remove surface debris with a bristle brush (NOT METAL BRISTLES) then wipe the surface with an alcohol soaked cloth.
9-6 Sterilization
Autoclave sterilization is recommended.
Autoclave sterilization is required for before the first use and then after each patient as noted below.
The following items can be autoclaved.
LED Handpiece with cord, Tip, Tip Wrench, Tip Holder, Handpiece Stand, Tube Holder, Sterilization Cassette.
1) Place the Handpiece and Tip Wrench in the Sterilization cassette. (Fig. 24)
2) Place the Tips onto the Tips Holder before putting the Tips Holder in the Sterilization Cassette.
3) Commence Autoclave sterilization. Autoclavable under the conditions below.
Autoclave for more than 20 min. at 121°C, or 15 min. at 132°C, or 3 min. at 134°C.
4) To maintain sterility, keep the product sealed in the autoclave pouch until required for use.
Tip Wrench
Tip Replacement
Wrench (for V10-S) and
VS Tip Wrench (Option)
Do not place Maintenance Tips in the Tip Holder, as these Tips are too long. Use an autoclave pouch when
sterilizing Maintenance Tips.
CAUTION
Do not autoclave the product with other instruments even when it is in a cassette. This is to prevent
possible discoloration and damage to the product from chemical residue from other instruments.
Keep the product in suitable atmospheric pressure, temperature, humidity, ventilation, and sunlight. The
LED Handpiece with cord
Sterilization Cassette locations for VarioSurg 3 components
16
Fiber end face
Cotton swab
Tip Holder
Tip holder alternative
location.
Tube Holder
Fig. 23
Fig. 24

Hide quick links:

Publicidad

Tabla de contenido

Solución de problemas

loading

Tabla de contenido