Application Programs; Weight Unit; Counting - Acculab SVI SERIE Manual De Instrucciones

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

ENGLISH
6
Please remove weight and zero scale.
7
Place cal-weight back on to pan and the result should be within +/- 1d.
8
Calibration complete.
Calibration Notes:
The actual calibration weight needs to be within +/- 5% of the selected weight.
If the scale is not zeroed before calibration, error "E-01" will be displayed.
If calibration weight is positively "above" the 5% tolerance, error "E-03" will be displayed.
If calibration weight is negatively "below" the 5% tolerance, error "E-03" will be displayed.
Always allow a minimum of thirty seconds for the scale to stabilize. During prolong periods of use
repeat calibration procedure at regular intervals for optimum accuracy.
Note: Please do not open or attempt to "internally" calibrate the scale. This will void warranty and may
damage scale .

Application Programs:

Weight Units: (Default Application)
1
To select "units" press and hold "TARE/ZERO" key for more than 3 seconds to access application
menu ("UNIT" will be displayed).
2
With "F" key you can scroll through the options (UNIT, COUNT, & TOTAL).
Select "UNIT" by pressing "CAL/PRINT" key (default unit "grams" or "kg" will be displayed).
3
4
With the "F" key you can scroll through weigh unit options (g, lbs, ozt & kg) depending upon model.
5
Select desired weigh unit by pressing "CAL/PRINT" (selected option will be displayed and balance
will return to weighing mode).
6
Pressing the "F" key will enable you to toggle between selected weight unit and default "kg".
7
Selected weigh unit will be retained in memory after balance is turned off.

Counting:

1
To select "Counting" application press and hold "TARE/ZERO" key for more than 3 seconds to access
application menu ("UNIT" will be displayed).
2
With "F" key you can scroll through the options (UNIT, COUNT, & TOTAL).
3
Select "COUNT" by pressing the "CAL/PRINT" key.
4
Display will automatically exit menu and return to weighing mode, with a "CNT" displayed in the left
hand side of the display & the selected weigh unit displayed (signifying to the user that they are in an
application mode).
5
Place empty container or box on scale and tare. (If applicable)
6
Press "F" key to display reference sample/default "1" (quickly pressing and releasing the "F" key will
advance the number 1,2,3, etc... / Holding the "F" key will advance the numbers faster 10, 20, 50, up
to 99).
7
After selection of the desired number, place corresponding samples on scale and press "CAL/PRINT"
to confirm (display will show "pcs" with corresponding piece count).
8
Pressing the "F" key will allow the user to toggle between gross weight, net weight, average piece
weight & pcs.
3