Oil priming

Before using your NSK iCare for the first time, and every time after the oil tank has been completely emptied, Follow these 8 steps to correctly
prime the oil tank pump. Failure to follow these instructions may result in your iCare not dispensing the correct oil volume.
Fill the NSK iCare oil tank with NSK iCare oil then:
1) Close the iCare door then, while holding down the Chuck Cleaning Button, switch ON (I side) the iCare main Power Switch.
2) A beep will then sound and the Error light will illuminate. Release the Chuck Cleaning Button.
3) Open the iCare door then hold down any Connector Select Button "1-4" and check that oil now flows from that selected Connector. Release
the Connector Select Button.
4) Close the iCare door.
5) Switch OFF (O side) the iCare main Power Switch. Now immediately switch ON (I side) the iCare main Power Switch.
6) Press each Connector Select Button so that all 4 Connector LED lights simultaneously illuminate.
7) Press the Start Button and oil priming will start. Oil should now start to flow from each Connector.
8) A BEEP will sound when the oil priming procedure has automatically finished.
7

Check before use

Before use, check the product by referring to the check points below. If you find something abnormal such as looseness, when checking or
using, stop use and contact your Authorized NSK Dealer.
• Check that the turbine connector part is securely fastened and the Nut screw is free of wear, seizure or adherence of dirt.
• Check that the E-type connector part is secure and free of flaws or adherence of dirt.
8

Operation Procedures

• Do not use iCare to lubricate air scalers (except for NSK Ti-Max S970 series) and Phatelus air motors. The iCare lubricating
system is not designed to lubricate those particular products. Maintain air scalers and Phatelus air motors as described in the
CAUTION
operation manual for each product.
• If blood infiltrates inside a handpiece, iCare may not totally clean the internal handpiece components, and internal blood
coagulation may cause product failure. In this case, NSK recommends to immediately lubricate those handpieces with PANA
SPRAY Plus / PANA SPRAY.
• Maintain individual handpiece according to the procedure described in the operation manual of handpiece.
8-1 Attaching Handpiece
Air Turbine
When connecting a Midwest 4 Hole air turbine (According to ISO 9168)
1) Remove the bur from the handpiece.
2) Open the iCare Door, insert the air turbine to the shape of the air turbine connector, and tighten the nut securely (Fig. 11).
To 4)
When connecting another type of air turbine
1) Remove the bur from the handpiece.
2) Open the iCare Door, insert the adaptor (option) to the shape of the air turbine connector, and tighten the nut securely (Fig. 12).
3) Insert the air turbine into the adaptor (Fig. 13).
* Refer to "16. Option Parts List" for the correct adaptor.
To 4)
4) Push and pull the handpiece to be sure that it is securely attached.
5) Point the bur insertion hole toward the door and close the door securely (Fig. 14).
E0643E002 (CS5) iCare-EN_141001.indd
8
8
2014/10/01
10:41:31