Technical specification
Nominal voltage
Nominal current
Protection class
Illuminance (without filters)
Colour temperature
Colour rendering index (CRI)
Cable length
Light field diameter at a
420 mm working distance
Vertical direction of illumination
Weight
Operation time
Environmental conditions
for operation
Environmental conditions
for storage
Environmental conditions
for transport
Device classification
according to IEC 62471
Consumables
LoupeLight 2 /
MicroLight 2
Loupes
Spare polarisation filter P2 for HR Loupes (2 pairs)
48/52
Sterilisable Swivel Lever for rigid loupe mount
(without i-View)
Sterilisable Swivel Lever for loupes with i-View
Yellow Filter
Polarisation filter P2
Connecting cord SC1 (2.4 mm Ø)
Connecting cord SC2 (3.2 mm Ø)
MED 113729
max. 350 mA
Internally powered (accu)
typ. 55,000 lx
typ. 4,000 K
typ. ≥ 90
approx. 1.50 m
approx. 80 mm
- 90° to ≥ 90°
- 20° to ≥ 20°
approx. 50 g (only light)
approx. 80 g (with S-Frame)
approx. 190 g (with Lightweight Headband)
typ. 4 h
(with mPack mini continuous operation)
+10 °C to +35 °C
30 % to 75 % rel. humidity
700 hPa to 1060 hPa
+5 °C to +45 °C
45 % to 80 % rel. humidity
500 hPa to 1060 hPa
-20 °C to +50 °C
45 % to 80 % rel. humidity
500 hPa to 1060 hPa
Group 2
2017-10-25
3 V DC
C-000.32.524
C-000.32.551
C-000.32.241
C-000.32.535
C-000.32.240
C-000.32.243
C-000.32.536